A quick reference to the optical formulas behind microscopy and microspectroscopy — the relationships that govern resolution, magnification and the quantitative spectral measurements CRAIC instruments perform. For spectral unit conversions (nanometers, microns and wavenumbers), see Units for Microspectroscopy.
Resolution and numerical aperture
- Numerical aperture: NA = n · sin θ — n is the refractive index of the medium between objective and sample; θ is the half-angle of the light cone the objective accepts.
- Abbe diffraction limit (lateral resolution): d = λ / (2 · NA)
- Rayleigh criterion: d = 0.61 · λ / NA
- Depth of field (approx.): DOF ≈ λ · n / NA2
Magnification
- Total magnification: Mtotal = Mobjective × Meyepiece (or × the camera projection factor)
Spectroscopy
- Transmittance: T = I / I0
- Absorbance (optical density): A = −log10(T)
- Beer–Lambert law: A = ε · c · l — absorbance is proportional to molar absorptivity ε, concentration c and path length l.
- Reflectance: R = Ireflected / Ireference
Please send suggestions for additional formulas to webmaster@craictechnologies.com.